Belimed Steam Sterilizers Belimed has been a leading provider of safe, reliable, efficient and easy to use sterilizers for over 40 years. The MST-V series of steam sterilizers is optimized for throughput and workflow efficiency. These sterilizers can be installed side-by-side, with no side or rear service access required, making lean, extremely space efficient department designs possible.
The MST-H series of carriage loading steam sterilizers offers three chamber sizes, each designed with a focus on high volume re-processing, process efficiency, and resource conservation.
In addition, steam-air-mixture processes for sterilization of liquids in closed containers are also possible. Saturated steam sterilizers for the typical sterilization of Bottles, Tubes, Pipe fittings, Metal containers, Containers, Cages, Textiles and Clothing, Syringes, Ampoules, Stoppers, Bags, Reagents and production items and waste and much more Autoclave freely adapted to the respective needs and tasks according to customer requirements. Key Information GMP compliant One and two-door versions Cabin capacity: from to litres Floor level loading possible High degree of individualisation options.
Dimensions: 21 different chamber sizes From 1 x 0. Suitable for. Downloads Brochure Technical Data sheets on request You haven't found what you're looking for?
Contact us here! Following maintenance, the number must be increased by the respective number of batches see Chap. Time Air Filter Reset Timer h 6. Date and time displayed in European or American standard format. Unit of Temp. Switch using buttons and see Chap. Substeri Decont1 6. Subsequently recalibrate doors see Chap. The DI boiler is temporarily kept up to temperature. The preheating time is only active outside of a program when the device is switched on.
It starts afresh after correct completion of a program cycle. If the switch code of the DI boiler is switched off, this function is deactivated see Chap. If 0 minutes is set, the DI boiler is emptied even if the target temperature has not been reached. The circulation pump is activated again upon program restart without pumping off.
If the first program step is M5 or M6, the entry is ignored. When a step is skipped see Chap. The first discharge of washing liquid must not take place before termination of the disinfection phase.
If the program is interrupted, the procedure must not be continued with discharge of the washing liquid before the disinfection phase has been completed. This prevents contaminated washing liquid from entering the waste water. In the event of an epidemic, the door interlock see Chap. On: Pumping off is always carried out in the event of program interruption normal state Off: Pumping off is not activated in the event of program interruption.
If a program is then started, the circulation pump starts subsequently without pumping off. Imported programs can be replaced using buttons 1 to 6.
Only: Program selection is possible only by means of rack coding. All sensors must close at least once when the rack is inserted, otherwise an error is displayed see Chap. Different programs can be activated, depending on the magnet combination. With button , the machine starts with specified program. Error Message Tech. The LS door can be opened after termination of the program.
Identification number must lie between barcode type EAN Identification name 20 characters max. The barcode scanner is also connected to Connector X A program can only be started once an entry has been made, otherwise the software remains on query. If "66" is entered, the text "No Identification" is forwarded.
After being entered, the number is forwarded to the printer and Batch Documentation System see Chap. Note: Only Digits can be used for a number. Two pre-rinse steps without comment appear in the batch documentation see Chap. If during repeat S opens again, an error signal is triggered, and pre-rinsing is discontinued with Error see Chap. Off If S opens during pre-rinsing, Error see Chap.
Caution All modifications of factory settings must be stored electronically or in paper form close to the device. This is only issued to Belimed-licensed service technicians after completing training for handling the controller. Operating Level Adjustment range 1.
It Drain Time is measured only after the empty level e. Residual water can lead to transfer of suds between water steps. See also Chap. Open door using button ton 2. Example: The value saved when exerting contact pressure is If the potentiometer value is while pressing together, Error "Door is open" is displayed see Chap. The lag time is added to the programmed dry- 30 sec ing time. Air and tank heaters are switched on when drying. Never set to "On" with electrical tank heater on!
The relay K28 is in the electro drawer. Connection can be made potential-free or with voltage, depending on the type of connection.
Output K28 Signals: Output K57 Program finished Error; program running Run ; end of program; open exhaust flap dry- Program run ing ; pre shut-off valves; chamber lighting Error Output K58 An error causes the current program to be interrupted and closes K28 , K57 and K Information for the house control system.
K28 closes before the relay K17 is activated. In automatically propelled exhaust flaps, the flap can be opened before the drying ventilator is activated.
Pre shut-off valve function K28 , K57 or K58 are closed when a program is running or the valve Y27 for DI water pre-heating is activated. Chamber lighting function K28 , K57 and K58 are closed when a program is running, the doors are open or a program has been finished in which the door has not yet been opened.
Delay interval is shown in milliseconds. Delay Rack Position ms Note: Rack must be positioned correctly. Reduce the delay interval to approx. Allocate Device Numbers Addresses are reserved for sterilisers. If communication fails for 10 s, Error appears on the display instead of "Program ready" see Chap. Digital Interface 1 Activate Error with "On". Year Manufacturing Date Year The batch num- Batch Counter Start ber must always be recorded beforehand and re-entered subse0 quently.
Every modification must be documented precisely. Caution Increasing the threshold values entails higher risk! The Belimed service technician is personally responsible for the modifications he has implemented.
The manufacturer accepts no responsibility for damages caused by operating errors or by not observing the instructions in the manual.
Last mod. Ao S6. Keep a print-out of the data with the device after taking the device into operation. Press button "Print operating" appears on the display. Activate printer using button Press button "Print operating data" appears on the display. During this display message, press button. Printing process starts Further information: See Chap. A new program cannot be activated. Temperature Sensor NTC4 is monitored.
Further information: See Chap. Type Abbreviation Description Pos. This applies to operational software of the device, and also with regard to databackup purposes. Clamp from left Data- 2. Adjusting and operating the: Printer see Chap. Barcode reader see Chap. Modem see Chap. After loading the new software, load the saved set up data to restore information such as machine serial number, manufacturing date, date of the last board change and batch count see Chap.
Turn off the main switch before carrying out any intervention on the device. Error causes can be systematically tracked with the analysis software see Chap. For errors which occur irregularly, the status of in- and outputs at the time the error occurred must be recorded via print-out see Chap. Poorly crimped or lacking connector contacts may be the cause.
Note Do not replace any components if the cause of the error has not been definitely clarified. Replaced parts may still be functioning. Label syntax e. Input S , R1 is monitored during the opening process. Input S , R1 is monitored during the closure process. If during the closure process the safety switch actuator opens Relay ESR1, the controller terminates the closure process with error display and reopens the door M Object in door Error Code Input S is monitored during the closure process.
Program is terminated with error display. If during the Object in door closure process the safety switch actuator S opens Relay ESR2, Error Code the controller terminates the closure process with error display and reopens with Motor M Input S , R2 is monitored during the closure process. If Switch No pressure S is still open during the program 90 s after the filling phase, the Error Code controller terminates the current program with error display.
Cause: Remedy: Wrong rotational direction of the circulation pump Change rotational direction, exchange Phases 1 and 2. Foam formation See Chap. If Switch SB in the floor pan closes for longer than 2 s, the controller terminates the current program with error display. Cause: Remedy: Coarse or fine screen is clogged Clean screens see Chap. If one of the switches SS and SS will not close, an error is displayed. When the LS door is open, all switches are monitored. When Switch S is closed, Switches SS , S and S must have been closed at least once beforehand, otherwise an error is displayed.
Cause: Remedy: Rack not correctly inserted Withdraw rack completely and insert again. No water Error Code If the level sensor P1 is not closed for longer than 2 s value between 0. The threshold values for Input P1 are monitored during the program cycle. The controller terminates Error Code the current program with error display see Chap. NTC 2 def. NTC 3 def. NTC 4 def. NTC 6 def.
No DI water supply to boiler see also Chap. Error Code If, in the filling process with CW Y11 , Operating Level 2 of the level sensor P1 is not reached within 5 minutes, the controller terminates the current program with error display see Chap. Threshold monitoring: Only during the filling process until Operating Level 2 is reached. Threshold adjustment for cold water see Chap. No warm water supply in the washing chamber, or switch S1, S2 open. No warm water Error Code If, in the filling process with CW Y12 , Operating Level 2 of the level sensor P1 is not reached within 5 minutes, the controller terminates the current program with error display see Chap.
Threshold monitoring: Threshold adjustment for warm water see Chap. No DI water If, in the filling process with DI water Y13 , Operating Level 2 of the Error Code level sensor P1 is not reached within 7 minutes, the controller terminates the current program with error display see Chap.
Threshold monitoring: Threshold adjustment for DI water supply see Chap. Error Code If Operating Level 2 is not reached within 3 minutes while the filling process with pre-heated DI water Y14 is under way, the controller terminates the current program with error display see Chap.
Error Code If the level does not drop to 0. Threshold monitoring: Only during drainage until empty level is reached. Drainage P1 3 min Threshold adjustment for drainage see Chap. If Switch S is in closed state while the drainage phase Y31 active is under way, the process is interrupted and an error is displayed. Threshold monitoring: 5 s after start of drainage.
Cause: Remedy: Switch S remains closed Replace Electric circuit between controller Repair interruption and valve interrupted Relay K31 does not respond Replacing the controller Drainage valve Y31 defective Check function by direct activation Y Clean or replace see Chap. Water lost in the washing phase. Level too low Error Code If the level drops below the minimum level of 1 Volt for 15 s in the course of the washing cycle, the controller terminates the current process with error display see Chap.
Threshold monitoring: 5 s after start of washing. Cause: Remedy: Drainage valve membrane soiled Clean or replace membrane see Chap. Steam is withdrawn Exhaust flap remains open stuck Repair exhaust flap Exhaust flap remains closed. Washing liquid is drained via Switch S31 see Chap. Unwanted entry of water into the washing chamber see Chap. Level too high Error Code If the level of 3. As long as the level is above 3. Threshold monitoring: As long as the device is switched on Cause: Remedy: Leaking water valves Replace water valves see Chap.
Flow Meter 1 does not detect required impulses. Wrong dosing D1 Error If a dosage has been programmed in the step phase with Flow Meter 1, at least 10 impulses of the flow meter must be counted within 1 minute during dosing with Motor M21 , and at least X impulses for every further 6 seconds.